Thermo Fischer Scientic RT7 troubleshooting

Repair NESLAB RTE7 waterbath

Thermo Fischer Scientific RTE7 troubleshooting

Thermo Fischer RTE7

Issue: ERROR 25  , display shows current temperature -99degres
Potential problems: dual element RTE temperature probe faulty : Thermo Fischer part numer : 082393RTE -> Price $120USD OR analog to digital converter port ( pin 9 or pin 10  faulty : ) U5 on the main board
or any elements going from the probe all the way to microcontroller.

FOUND : resistor R81 ( value : 1.21kOhm ) opened  - reads 3.3Mega ohms instead.
replacing this resistor completes the repair.




attached pictures and dual temperature sensor DIAGRAM for NESLAB RTE7






Main microcontroller RTE7
Analog to Digital converter beside microcontroller 
LT1490A - part of the temperature sensing circuit
Thermo Fischer RTE7 model and serial number
GRAY and ORANGE for dual temp sensor
top of the main board of RTE7 - by Thermo Fischer
 

No comments: