Repair NESLAB RTE7 waterbath
Thermo Fischer Scientific RTE7 troubleshooting
Thermo Fischer RTE7 |
Issue: ERROR 25 , display shows current temperature -99degres
Potential problems: dual element RTE temperature probe faulty : Thermo Fischer part numer : 082393RTE -> Price $120USD OR analog to digital converter port ( pin 9 or pin 10 faulty : ) U5 on the main board
or any elements going from the probe all the way to microcontroller.
FOUND : resistor R81 ( value : 1.21kOhm ) opened - reads 3.3Mega ohms instead.
replacing this resistor completes the repair.
attached pictures and dual temperature sensor DIAGRAM for NESLAB RTE7
Main microcontroller RTE7 |
Analog to Digital converter beside microcontroller |
LT1490A - part of the temperature sensing circuit |
Thermo Fischer RTE7 model and serial number |
GRAY and ORANGE for dual temp sensor |
top of the main board of RTE7 - by Thermo Fischer |
No comments: